Analytical HPLC System LAHS-A10

Analytical HPLC System LAHS-A10 features Quaternary Low Pressure Pump with 0.001-10 ml/min pump flow rate. Incorporated UV-Vis detector with a range of 190 to 800 nm that examines sample absorption at different wavelengths and a high pressure pump providing continuous stable flow. This HPLC system offers high speed accuracy, resolving power, detection, quantification, convenience and applicability to new sample types. 

Buy Now View Catalog
Pump specification
  • Type: Quaternary Low Pressure Pump
  • Flow Rate Range: 0.001 to 10.000 ml/min, in 0.001ml increments
  • Flow Accuracy: ≤ ± 0.2% (1.00 ml/min, H2 O, R.T., 8.5 MPa)
  • Flow Precision: RSD ≤ 0.06% (1.00 ml/min, H2 O, R.T., 8.5MPa)
  • Maximum Pressure: 63 MPa
  • Pressure accuracy: ≤ ± 3.0 %
  • Pressure pulsation: ≤ 0.1 MPa
  • Composition rang: 0 to 100 %
  • Composition precision: ≤ 0.3 % RSD
UV/VIS Detector
  • Wavelength Range: 190 to 800 nm
  • Wavelength Repeatability: ≤ 0.1 nm
  • Wavelength Accuracy: ≤ ± 1.0 nm
  • Band Width: 6 nm
  • Response Time: 0.0 s to 5 s
  • Baseline Noise: ≤ ± 0.25 × 10-5 AU(empty cell, 254 nm, 1.0s)
  • Drift: ≤ 0.5 × 10-4 AU/h (empty cell, 254 nm, 60 min)
  • Path length: 10 mm(analytical cell)
  • Light Source: Deuterium Lamp and Tungsten Lamp
Data-processing Workstation System contain the tables for regular samples
Equipped with Internal and external standard calculation process
Automatic display, print, export of data and starting of another program on completion on measurement
Observe with the condition of 21 CFR part 11 directive of the FDA
System Kits 1 Set
Combined A/D converters
Combined vacuum degasser and 4 channel
Power Supply AC 110 /220 V ±1 0 %, 50/60 Hz
Dimension 710 × 583× 500mm; 770 × 450 × 700 mm
Weight 156 kg
  • UV-Vis detector with a range of 190 to 800 nm
    Consist of Quaternary Low Pressure Pump
    Advance and intellectual HPLC system
    Excellent Flow Accuracy and Pressure Accuracy
    Column oven with greater capacity of 3 column with maximum thickness ID10 × 250 mm
    Fast and stable UDP communication protocol, can be used for instrument remote control
    Enhancement of up to 5 mm stroke of plunger rod with good performance and high pressure constant flow pump or parallel pump head
    Complies with FDA 21 CFR part 11
Degasser
Accessories no. Name
1 Column oven
2 Auto sampler
3 Degasser
4 Diode Array Detector
5 Refractive Index Detector
6 Evaporative Light Scattering Detector
7 Fluorescence Detector

High Performance Liquid Chromatography system has been widely used in Research and Development, pharmaceutical, food nutrient analysis of amino acids and sugars, detection of food contaminants, environmental pollutant and chemical product purity.

{"Pump specification":{"Type":"Quaternary Low Pressure Pump","Flow Rate Range":"0.001 to 10.000 ml\/min, in 0.001ml increments","Flow Accuracy":"\u2264 \u00b1 0.2% (1.00 ml\/min, H2 <\/sub>O, R.T., 8.5 MPa)","Flow Precision":"RSD \u2264 0.06% (1.00 ml\/min, H2 <\/sub>O, R.T., 8.5MPa)","Maximum Pressure":"63 MPa","Pressure accuracy":"\u2264 \u00b1 3.0 %","Pressure pulsation":"\u2264 0.1 MPa","Composition rang":"0 to 100 %","Composition precision":"\u2264 0.3 % RSD"},"UV\/VIS Detector":{"Wavelength Range":"190 to 800 nm","Wavelength Repeatability":"\u2264 0.1 nm","Wavelength Accuracy":"\u2264 \u00b1 1.0 nm","Band Width":"6 nm","Response Time":"0.0 s to 5 s","Baseline Noise":"\u2264 \u00b1 0.25 \u00d7 10-5 <\/sup> AU(empty cell, 254 nm, 1.0s)","Drift":"\u2264 0.5 \u00d7 10-4 <\/sup> AU\/h (empty cell, 254 nm, 60 min)","Path length":"10 mm(analytical cell)","Light Source":"Deuterium Lamp and Tungsten Lamp"},"Data-processing Workstation":"System contain the tables for regular samples\nEquipped with Internal and external standard calculation process\nAutomatic display, print, export of data and starting of another program on completion on measurement\nObserve with the condition of 21 CFR part 11 directive of the FDA","System Kits":"1 Set\nCombined A\/D converters\nCombined vacuum degasser and 4 channel","Power Supply":"AC 110 \/220 V \u00b11 0 %, 50\/60 Hz","Dimension":"710 \u00d7 583\u00d7 500mm; 770 \u00d7 450 \u00d7 700 mm","Weight":"156 kg"}
UV-Vis detector with a range of 190 to 800 nm
Consist of Quaternary Low Pressure Pump
Advance and intellectual HPLC system
Excellent Flow Accuracy and Pressure Accuracy
Column oven with greater capacity of 3 column with maximum thickness ID10 × 250 mm
Fast and stable UDP communication protocol, can be used for instrument remote control
Enhancement of up to 5 mm stroke of plunger rod with good performance and high pressure constant flow pump or parallel pump head
Complies with FDA 21 CFR part 11
High Performance Liquid Chromatography system has been widely used in Research and Development, pharmaceutical, food nutrient analysis of amino acids and sugars, detection of food contaminants, environmental pollutant and chemical product purity.


Related Products of Analytical HPLC

Analytical HPLC System LAHS-B10

Analytical HPLC System LAHS-B10

High Pressure Constant Flow Pump :

  • Double plunger reciprocating pump: 1 set
  • Flow range: 0.001 to 10.00 mL/min
  • Setting step size: 0.001 mL/min
  • Flow accuracy: ≤ ± 0.3 %
  • Flow stability: RSD ≤ 00.1 %
  • Maximum pressure: 42 MPa
  • Display pressure error: ≤ 1.0 %
  • Pressure fluctuation: ≤ 1.0 %
  • Power consumption: 80 W

UV/Visible Detector :

  • Wavelength range: 190 to 700 nm
  • Wavelength repeatability: ± 0.1 nm
  • Band width: 8 nm
  • Response time: 0 to 4.9 seconds
  • Linearity range: 1.8 AU (5 %)
  • Baseline Noise: ≤ ± 1 × 10-5AU
  • Baseline drift: ≤ 2.0 × 10-4AU/h
  • Minimum detectable concentration: 4.0 × 10-9 g/mL
  • Detector cell: Optical path: 10 mm
    Volume: 10 µL
  • Light source: Deuterium lamp
  • Power consumption: 100 W
  • Voltage: AC 110 V/ 220 V ± 10 %, 50/60 Hz
  • Packing Dimensions (W × D × H): 600 × 440 × 380 mm
    2 × (600 × 410 × 300 mm
  • Gross weight: 54 kg

Analytical HPLC System LAHS-B11

Analytical HPLC System LAHS-B11

High Pressure Constant Flow Pump :

  • Double plunger reciprocating pump: 2 sets
  • Flow range: 0.001 to 10.00 mL/min
  • Setting step size: 0.001 mL/min
  • Flow accuracy: ≤ ± 0.3 %
  • Flow stability: RSD ≤ 00.1 %
  • Maximum pressure: 42 MPa
  • Display pressure error: ≤ 1.0 %
  • Pressure fluctuation: ≤ 1.0 %
  • Power consumption: 80 W

UV/Visible Detector :

  • Wavelength range: 190 to 700 nm
  • Wavelength repeatability: ± 0.1 nm
  • Band width: 8 nm
  • Response time: 0 to 4.9 seconds
  • Linearity range: 1.8 AU (5 %)
  • Baseline Noise: ≤ ± 1 × 10-5AU
  • Baseline drift: ≤ 2.0 × 10-4AU/h
  • Minimum detectable concentration: 4.0 × 10-9 g/mL
  • Detector cell: Optical path: 10 mm
    Volume: 10 µL
  • Light source: Deuterium lamp
  • Power consumption: 100 W
  • Voltage: AC 110 V/ 220 V ± 10 %, 50/60 Hz
  • Packing Dimensions (W × D × H): 600 × 440 × 380 mm
    3 × (600 × 410 × 300 mm
  • Gross weight: 75 kg

Analytical HPLC-A7840
Analytical HPLC-A7840

Pump Specifications :

  • Solvent delivery: Double plunger series mode, floating plunger design, single stroke volume 30 µl
  • Pump Head: 10 ml stainless steel pump head
  • Flow rate range: 0.001 to 9.999 ml / min
  • Flow rate increment: 0.001 ml/min
  • Maximum pressure: 42 MPa
  • Column size: 4.6 × 250 mm 5 µm
  • Pipeline Size: 1/16 inch
  • Flow rate accuracy: ± 0.5 %
  • Flow rate precision: RSD ≤ 0.1%
  • Gradient accuracy %: ± 0.5 %
  • Gradient repeatability %: RSD ≤ 0.15%
  • GLP: Automatic recording of pump sealing ring working time
  • Display: LCD backlight display
  • Power: 80 W
  • Power supply: 220 V ± 10%, 50 Hz
  • Dimension: 398 × 149 × 267 mm

Detector Specifications :

  • Wavelength range: 190 to 700 nm
  • Wavelength accuracy: ± 1 nm
  • Wavelength repeatability: 0.2 nm
  • Deuterium lamp: 190 to 400 nm
  • Tungsten lamp: 400 to 700 nm
  • Baseline noise ( Static ) Au: 1 × 10-5 AU/h at 254 nm with methanol/water= 80/20
  • Baseline drift ( Static ) Au / h: 1.5 × 10-4 AU/h at 254 nm with methanol/ water= 80/20
  • Time constants: 16 grades adjustable
  • Display: LCD backlight display
  • GLP: Total number of lights, lamp replacement time and product serial number
  • Power: 75 W
  • Power supply: 220 V ± 10%, 50 Hz
  • Dimension (L × W × H): 398 × 149 × 267 mm

Analytical HPLC-A7832
Analytical HPLC-A7832

Detector Specifications :

Model No HPD-A20 HPD-A21
Wavelength range Single wavelength ( 190 to 800 nm ) Dual wavelength ( 190 to 800 nm )
Wavelength accuracy ± 0.5 nm ± 2 nm
Wavelength repeatability 0.15 nm 0.3 nm
Band width 8 nm 8 nm
Optical path 10 nm 10 nm
Deuterium lamp 190 to 400 nm 190 to 400 nm
Tungsten lamp 400 to 800 nm 400 to 800 nm
Baseline noise ( Static ) Au ≤ ± 0.35 × 10-5 ≤ ± 0.5 × 10-5
Baseline drift ( Static ) Au / h ≤ 1 × 10-4 ≤ 1.5 × 10-4
Minimum detection limit ≤ 4 × 10-8 ≤ 4 × 10-8
Power 100 W 100 W
Power supply 90 V / 50 to 60 Hz 90 V / 50 to 60 Hz
Dimension 460 × 315 × 180 mm 460 × 315 × 180 mm
Weight 9.5 kg 9.5 kg

Pump Specifications :

  • Flow rate range: 0.0001 to 10 ml / min
  • Maximum pressure: 50 MPa , 0 to 2 ml / min
  • Column size: 4.6 × 250 mm
  • Sample size: 5 µm
  • Pressure pulsation: ≤ 1 % at 10 MPa
  • Flow rate accuracy: ± 0.15 % at 1 ml / min
  • Flow rate repeatability %: ≤ 0.1
  • Gradient accuracy %: ± 1.5
  • Gradient repeatability %: ≤ 0.2
  • Dimension: 460 × 310 × 180 mm
  • Power supply: 90 V
  • Weight: 11.6 kg

Analytical HPLC-A7835

Analytical HPLC-A7835

Pump Specifications :

  • Model No: HPP-B1
  • Flow rate range: 0.0001 to 10 ml / min
  • Maximum pressure: 42 MPa , 0 to 5 ml / min
  • Column size: 4.6 × 250 mm
  • Sample size: 5 μm
  • Pressure pulsation: ≤ 1.5 % at 10 MPa
  • Flow rate accuracy: ± 0.2 % at 1 ml / min
  • Flow rate repeatability %: ≤ 0.15
  • Gradient accuracy %: ± 2
  • Gradient repeatability %: ≤ 0.2
  • Dimension: 450 × 260 × 190 mm
  • Power supply: 90 V to 240 V
  • Weight: 11.2 kg

Detector Specifications :

  • Model No: HPD-B2
  • Wavelength range: 190 to 800 nm
  • Wavelength accuracy: ± 0.75 nm
  • Wavelength repeatability: 0.2 nm
  • Band width: 8 nm
  • Optical path: 10 nm
  • Deuterium lamp: 190 to 400 nm
  • Tungsten lamp: 400 to 800 nm
  • Baseline noise ( Static ) Au: ≤ ± 0.5 × 10 -5
  • Baseline drift ( Static ) Au / h: ≤ 1.5 x 10 -4
  • Minimum detection limit: ≤ 4 x 10 -8
  • Power: 100 W
  • Power supply: 90 V / 50 to 60 Hz
  • Dimension: 460 × 260 × 190 mm
  • Weight: 9.2 kg

Standard HPLC

HPLC system LHLC-A10

High pressure pump Single pump
Flow rate range 0.001 - 9.999 mL / min (0.001 mL / min increment)
Pressure range ≤ 0.5 MPa (0 - 42 MPa)

HPLC system LHLC-A11

High pressure pump Two sets of pump
Flow rate range 0.001 - 9.999 mL / min (0.001 mL / min increments)
Pressure range ≤ 0.5 MPa (0 - 42 MPa)